Arrhythmias & electrophysiologyCohort study

Long-term performance of single-connector (DF4) implantable defibrillator leads

EuropacePublished

  • Retrospective cohort study

Authors' conclusion

DF4 defibrillator leads demonstrate excellent longevity with >98.3% of leads followed for at least 5 years still functioning normally. Younger age at implant and lead manufacturer are associated with an increased risk of DF4 lead malfunction.

Study context

Population
Patients with single-connector (DF4) ICD leads, 2011–2023
Participants
5,289 patients
Outcome
Lead malfunction (electrical abnormalities requiring revision)

Main finding

As reported in the abstract

mean follow-up 4.15 ± 3.6 years

Leads with electrical abnormalities requiring revision

80 (1.5%)

Average time to failure (years)

4 ± 2.8 (median = 3.5)

Observational data: associations, not effects.

Further findings

Management of malfunctioned leads

Extracted and replaced
62/80 (78%)
Abandoned, new lead implanted
18

Predictors of malfunction (multivariable)

Younger age at implant, OR per year
1.03; P < 0.001
Abbott/St. Jude leads increased risk of malfunction

Original publication

Long-term performance of single-connector (DF4) implantable defibrillator leads. Europace. Published 24 November 2023. doi:10.1093/europace/euad347

Summarised from the published abstract; every number is checked against it