Arrhythmias & electrophysiology

Epicardial placement of implantable cardioverter-defibrillators in adults: Technical considerations, system performance, and clinical outcomes

Heart RhythmPublished Case series

Authors' conclusion

Surgical implantation of epicardial ICD systems with HV coils/leads is safe and feasible.

Study context

Population
Adults receiving surgically implanted epicardial implantable cardioverter-defibrillator (ICD) high voltage (HV) coils or leads, Hospital of the University of Pennsylvania, 2014–202514 men, median age 64 years
Participants
19 adults
Outcome
Indications, techniques, safety and long-term performance
Follow-up
Median 468 days
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Indication: ICD replacement after lead extraction for endocarditis
7
Indication: placement during concomitant surgery, incl. for tricuspid regurgitation
6
Indication: LV assist device
3
Indication: lack of venous access
3
HV coils used
58%
Epicardial leads received
42%
CRT received
68%
HV coil/lead site: posterior LV
48%
HV coil/lead site: anterior RV
16%
HV coil/lead site: both ventricles
26%
HV coil/lead site: transverse sinus
11%

Observational data: associations, not effects.

Further findings

System performance and mortality · median follow-up 468 days

Impedances > 200 Ω (loose set screw; damaged connector pin)
2
Impedances < 20 Ω without clinical impact
2
Coil fracture 1.5 years post-implant
1
Failed shocks or lead migrations
0
Died, largely from advanced HF; none attributable to surgery
26%

Original publication

Epicardial placement of implantable cardioverter-defibrillators in adults: Technical considerations, system performance, and clinical outcomes. Heart Rhythm. Published 22 May 2025. doi:10.1016/j.hrthm.2025.05.034

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.