Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time

Circ Cardiovasc ImagingPublished Multicentre reproducibility study

Authors' conclusion

The average bias-MBF and bias-MFR values were small with standard deviations substantially less than the test-retest variability. This demonstrates that MBF can be measured consistently across multiple sites and further supports that this technique can be reliably implemented.

Study context

Population
Patients with intermediate to high pretest probability of CAD undergoing rest and pharmacological stress Tc-99m-tetrofosmin CZT SPECT myocardial blood flow (MBF) studies at 5 sites (Canada, Italy, Japan, Germany, Singapore)25 to 35 studies per site
Comparison
Local-site vs core-laboratory SPECT MBF processing
Outcome
Intersite processing repeatability of SPECT MBF and additional camera time required
Methodological details5 sites, core laboratory comparison
  • 5 sites
  • Core laboratory comparison

Main finding

As reported in the abstract

Correlation, MBF at rest
0.93 (range, 0.87-0.97)
Correlation, MBF at stress
0.90 (range, 0.84-0.96)
Correlation, myocardial flow reserve
0.84 (range, 0.70-0.92)
Bias-MBF at rest, median (IQR)
5.4% (−3.8% to 14.8%)
Bias-MBF at stress, median (IQR)
5.4% (−6.2% to 19.4%)
Bias-MBF range between sites at rest
−1.6% to 11.0%
Bias-MBF range between sites at stress
−1.9% to 16.3%
IQR in bias-MBF between sites at rest
between 9.3% (4.8%-14.0%) and 22.3% (−10.3% to 12.0%)
IQR in bias-MBF between sites at stress
between 17.0% (−11.3% to 5.6%) and 33.3% (−10.4% to 22.9%)
Overall difference in myocardial flow reserve
1.52% (−10.6% to 11.3%)
Additional acquisition time, average, range between sites
44 to 79 minutes

Further findings

No significant differences between with and without motion correction

Original publication

Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time. Circ Cardiovasc Imaging. Published 6 October 2023. doi:10.1161/circimaging.122.015009

Composed by AI from the published abstract and checked against it. How Briefs are made →