New-Onset Conductance Disturbances After Transcatheter Tricuspid Valve Replacement: A Mechanistic Assessment

JACC Cardiovasc IntervPublished Retrospective cohort study

Authors' conclusion

NOCD incidence was 44.6% after TTVR in a highly selected patient population. Baseline hyperdynamic right ventricular function and contact of the device with the MS were independently associated with NOCD at 30 days.

Study context

Population
Patients undergoing transcatheter tricuspid valve replacement (TTVR)
Participants
70 patients
Exposure
New-onset conductance disturbance (NOCD)+ vs NOCD- after TTVR
  • NOCD+n = 31
Outcome
NOCD incidence
Follow-up
30 days
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Developed NOCD (RBBB, complete AV block, slow AF)

31

44.3%

Observational data: associations, not effects.

Further findings

Type IIIB leaflet morphology

P = 0.049

Postprocedural white blood cell count peak >13 × 109/L

P = 0.046

OutcomeNOCD+NOCD-P value
Baseline absolute RV free wall longitudinal strain29.7% ± 5.4%25.1% ± 6.4%P = 0.002

New permanent pacemaker implantation · 30 days

New permanent pacemaker
8 patients (14%)

Membranous septum dimensions, device oversizing

No difference

Predictors of NOCD (logistic regression)

Associated with baseline absolute RV free wall longitudinal strain ≥29% and subvalvular device contact with the membranous septum (MS) on follow-up CT

Original publication

New-Onset Conductance Disturbances After Transcatheter Tricuspid Valve Replacement: A Mechanistic Assessment. JACC Cardiovasc Interv. Published 14 October 2025. doi:10.1016/j.jcin.2025.07.039

Summarised from the published abstract; every number is checked against it