Arrhythmias & electrophysiology

Outcomes of leadless pacemaker implantation following transvenous lead extraction in high-volume referral centers: Real-world data from a large international registry

Heart RhythmPublished Multicentre registry cohort study

Authors' conclusion

LPMs showed a satisfactory safety and efficacy profile after TLE. Better electrical parameters were obtained when LPMs were implanted at a different RV location than the one where the previous transvenous RV lead was extracted.

Study context

Population
Consecutive patients undergoing leadless pacemaker implantation, International LEAdless PacemakEr Registry15.6% previous TLE
Participants
1,179 patients
Exposure
LPM implantation after TLE vs de novo LPM implantation
Outcome
LPM-related complication rate at implantation and during FU
Follow-up
Median FU 33 months (IQR 24-47)
Methodological details12 European centres, consecutive patients
  • 12 European centres
  • Consecutive patients

Main finding

As reported in the abstract

LPM-related major complications · median FU 33 months

P = 0.785

Observational data: associations, not effects.

Secondary outcomes

All-cause mortality · median FU 33 months

P = 0.288

Very high pacing threshold (>2 V@0.24 ms)

P = 0.004

High pacing threshold (>1-2 V@0.24 ms), LPM at different right ventricular (RV) location vs same location as extracted lead (TLE subgroup) · implantation, 1-month follow-up (FU), 12-month FU · implantation: 5.9% vs 18.2%
P = 0.012
High pacing threshold (>1-2 V@0.24 ms), LPM at different RV location vs same location as extracted lead (TLE subgroup) · implantation, 1-month FU, 12-month FU · 1-month FU: 3.4% vs 12.9%
P = 0.026
High pacing threshold (>1-2 V@0.24 ms), LPM at different RV location vs same location as extracted lead (TLE subgroup) · implantation, 1-month FU, 12-month FU · 12-month FU: 4.3% vs 14.5%
P = 0.037

Very high pacing threshold (>2 V@0.24 ms)

Pacing threshold higher in TLE group at implantation and during FU

Safety

LPM-related major complications · median FU 33 months

P = 0.785

Original publication

Outcomes of leadless pacemaker implantation following transvenous lead extraction in high-volume referral centers: Real-world data from a large international registry. Heart Rhythm. Published 7 December 2022. doi:10.1016/j.hrthm.2022.12.002

Composed by AI from the published abstract and checked against it. How Briefs are made →