Prediction of fractional flow reserve using intravascular ultrasound
Authors' conclusion
The IQ-FFR is a promising method to identify coronary lesions requiring revascularisation and to predict the FFR after stent implantation.
Study context
- Population
- Coronary lesions with stenosis (25–99%) and intravascular ultrasound (IVUS) and fractional flow reserve (FFR) measured just before and after stent implantation
- Participants
- 270 coronary lesions
- Comparison
- IVUS-derived IQ-FFR vs wire-derived FFR
- Outcome
- Agreement of IQ-FFR with wire-derived FFR
Methodological detailsprospective, single-centre
- Prospective
- Single-centre
Performance
As reported in the abstract
| Test or model | Correlation with wire-derived FFR (r) | Accuracy predicting post-stent FFR >/<0.80 |
|---|---|---|
| IQ-FFR | 0.896 | 87.5% |
IQ-FFR
Correlation with wire-derived FFR (r)0.896
Accuracy predicting post-stent FFR >/<0.8087.5%
Original publication
Prediction of fractional flow reserve using intravascular ultrasound. EuroIntervention. Published 7 October 2024. doi:10. 4244/ eij-d-24-00010
Composed by AI from the published abstract and checked against it. How Briefs are made →