Authors' conclusion

The inspIRE study confirms the safety and effectiveness of pulmonary vein isolation using the novel 3-dimensional mapping integrated circular loop catheter. An optimal number of PFA applications (≥48 total or ≥12 per vein) resulted in an improved 1-year success rate of ≈80%.

Study context

Population
Drug-refractory paroxysmal AF
Participants
186 participants
Outcome
Acute pulmonary vein isolation plus freedom from any atrial arrhythmia
Follow-up
12 months
Methodological detailsprospective, multicentre, Follow-up: 24-hour Holter at 3, 6 and 12 months plus remote rhythm monitoring (weekly 3–5 months, monthly 6–12 months, and for symptoms)
  • Prospective
  • Multicentre
  • Follow-up: 24-hour Holter at 3, 6 and 12 months plus remote rhythm monitoring (weekly 3–5 months, monthly 6–12 months, and for symptoms)

Main finding

As reported in the abstract

PEE (acute PVI plus freedom from any atrial arrhythmia)

PEE achieved

75.6%

95% CI 69.5%–81.8%

Observational data: associations, not effects.

Further findings

Freedom from symptomatic recurrence (clinical success)

Freedom from symptomatic recurrence
81.7% (95% CI, 76.1%-87.2%)

Freedom from all and symptomatic recurrence, simulated standard real-world monitoring (no protocol-driven remote monitoring)

Freedom from all recurrence
85.8% (95% CI, 80.8%-90.9%)
Freedom from symptomatic recurrence
94.0% (95% CI, 90.6%-97.5%)

Independent predictors of PEE success (multivariate)

LVEF ≥60%, adjusted OR
0.30
≥48 PFA applications, adjusted OR
0.28

PEE success by PFA applications per vein

≥12 PFA applications per vein
79.2%
Fewer PFA applications
57.1%

Safety

Primary adverse event rate

Primary adverse event rate
0%
  • Previously reported

Serious procedure-related AE

Serious procedure-related AE (urinary retention)
1

Original publication

Predictors of Success for Pulmonary Vein Isolation With Pulsed-field Ablation Using a Variable-loop Catheter With 3D Mapping Integration: Complete 12-month Outcomes From inspIRE. Circ Arrhythm Electrophysiol. Published 24 April 2024. doi:10.1161/circep.123.012667

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.