Arrhythmias & electrophysiologyResearch article

Quantitative Vector Screening to Improve Sensing and Reduce Inappropriate Shocks With the Subcutaneous Implantable Cardioverter Defibrillator

Circ Arrhythm ElectrophysiolPublished

  • consecutive patients who underwent subcutaneous implantable cardioverter defibrillator (S-ICD) implantation
  • single centre (Hospital of the University of Pennsylvania)
  • before–after comparison of screening protocols

Authors' conclusion

Implementation of a novel S-ICD screening protocol with stricter eligibility thresholds reduced sensing-related complications, particularly inappropriate shocks.

Study context

Population
Patients undergoing S-ICD implantation at the Hospital of the University of Pennsylvania from December 2018 to July 2025consecutive patients who underwent S-ICD implantation · traditional vector screening arm 145 · quantitative vector screening (QVS) arm 78
Participants
223 patients
Comparison
QVS protocol vs traditional vector screening
  • QVSn = 78
  • traditional vector screeningn = 145
Outcome
Time to first inappropriate shock or under-sensed ventricular arrhythmia
Follow-up
Median 42 months (traditional) and 18 months (QVS)

Main finding

As reported in the abstract

P = 0.02

Secondary outcomes

Patient eligibility during preimplant screening

Follow-up after implant

Original publication

Quantitative Vector Screening to Improve Sensing and Reduce Inappropriate Shocks With the Subcutaneous Implantable Cardioverter Defibrillator. Circ Arrhythm Electrophysiol. Published 20 April 2026. doi:10.1161/circep.125.014581

Summarised from the published abstract; every number is checked against it