Valvular & structural heart diseaseCohort study

Risk of Aortic Root Rupture in Bicuspid Aortic Stenosis: The ABC Sizing Algorithm for SAPIEN 3 Valves

Circ Cardiovasc IntervPublished

  • retrospective multicenter study
  • 15 centers in 7 countries
  • case-control (aortic root rupture cases and controls without root complications)
  • computed tomography analysis blinded to rupture status

Authors' conclusion

High calcium volume, narrow sinus dimensions, and excessive valve oversizing were associated with aortic root rupture. The ABC algorithm showed promising performance for risk stratification of patients with bicuspid anatomy being considered for treatment with a SAPIEN 3 valve.

Study context

Population
Patients with bicuspid aortic stenosis treated with transcatheter aortic valve replacement using SAPIEN 3 valves23 ruptures and 147 controls · ruptures 23 · controls 147
Participants
170 patients
Exposure
ABC Bicuspid Sizing Algorithm for the identification of rupture cases
  • Casesn = 23
  • Controlsn = 147

Main finding

As reported in the abstract

Total calcium volume

P < 0.001

Observational data: associations, not effects.

Further findings

Maximum sinus difference · smaller in cases

P < 0.001

OutcomeCasesControlsP value
Intercommissural distance difference · smaller in casesmedian −0.15 mmmedian 1.19 mmP = 0.009
Annular oversizing · greater in casesmedian 6.57%median 3.51%P = 0.047

Diagnostic performance of any high-risk criterion for identification of rupture cases

Sensitivity
100% (95% CI, 85.7%-100%)
Specificity
89.1% (95% CI, 83.1%-93.2%)
Diagnostic odds ratio
374.5 (95% CI, 21.7-6459.9)
Positive likelihood ratio
9.19
Negative likelihood ratio
0.023

Original publication

Risk of Aortic Root Rupture in Bicuspid Aortic Stenosis: The ABC Sizing Algorithm for SAPIEN 3 Valves. Circ Cardiovasc Interv. Published 25 June 2026. doi:10.1161/circinterventions.126.016918

Summarised from the published abstract; every number is checked against it