Valvular & structural heart disease

Redefining Transcatheter Aortic Valve Replacement (TAVR) Valve Sizing: A Validated Multiplanar Approach for Both Bicuspid and Tricuspid Valves

JACC Cardiovasc IntervPublished Retrospective cohort study

Authors' conclusion

Multiplanar supra-annular measurement provides a comprehensive evaluation of aortic root anatomy, enabling more precise TAVR sizing for both BAV and TAV patients.

Study context

Population
Patients with bicuspid aortic valve (BAV) or TAV undergoing TAVR with the Venus A-Valve191 BAV, 186 TAV
Participants
377 patients
Exposure
Multiplanar supra-annular sizing vs traditional annulus-based sizing
Methodological detailspropensity-matched comparison with VENUS-A cohort
  • Propensity-matched comparison with VENUS-A cohort

Main finding

As reported in the abstract

Downsizing vs annulus-based sizing

Observational data: associations, not effects.

Further findings

OutcomeMultiplanar strategyAnnulus-basedOR (95% CI)P value
Valve-in-valve procedures3.1%12.2%3.9 (1.206–16.186)P = 0.032
  • Valve-in-valve procedures: propensity-matched cohort (n = 98 pairs)

Postoperative morphologic and biomechanical analyses (subset n = 54)

Strong correlations between preprocedural measurements and postimplantation valve morphology and biomechanics

Original publication

Redefining TAVR Valve Sizing: A Validated Multiplanar Approach for Both Bicuspid and Tricuspid Valves. JACC Cardiovasc Interv. Published 28 July 2025. doi:10.1016/j.jcin.2025.05.043

Summarised from the published abstract; every number is checked against it