Arrhythmias & electrophysiology

SMART Pass automatic deactivation of subcutaneous implantable cardioverter-defibrillators

Heart RhythmPublished Retrospective cohort study

Authors' conclusion

The rate of inappropriate shocks was numerically higher after SMART Pass deactivation. A low QRS amplitude in S-ECG was a potential risk factor for SMART Pass deactivation.

Study context

Population
Patients who underwent subcutaneous ICD implantation between 2016 and 2021
Participants
115 patients
Exposure
SMART Pass deactivation vs SMART Pass consistently on
  • SP-OFFn = 12
  • SP-ONn = 87reference
Outcome
Inappropriate shocks
Follow-up
Median 1094 (IQR 887-1502) days

Main finding

As reported in the abstract

SP-OFF

3

events

SP-ON

4

events

Observational data: associations, not effects.

Further findings

OutcomeEstimateP value
S-ECG QRS amplitude, pre vs at SMART Pass deactivationSignificant temporary drop led to SMART Pass deactivation(95% CI 0.3–1.9)P = 0.015

Pre- and post-S-ECG QRS amplitudes

Significantly lower in SP-OFF than SP-ON
both P < .05
Similar 12-lead electrocardiogram (ECG) QRS amplitude in both groups

Original publication

SMART Pass automatic deactivation of subcutaneous implantable cardioverter-defibrillators. Heart Rhythm. Published 23 August 2024. doi:10.1016/j.hrthm.2024.08.037