Stent frame deformation of self-expanding transcatheter heart valves in bicuspid aortic stenosis and impact on valve performance
Authors' conclusion
The self-expanding Evolut™ THV demonstrates consistent valve performance in bicuspid AS. Stent frame under-expansion and ellipticity at the inflow level do not compromise valve haemodynamics or contribute to leaflet thickening at short-term follow-up.
Study context
- Population
- Patients with bicuspid aortic stenosis (AS) undergoing transcatheter aortic valve implantation (TAVI) with the self-expanding Evolut transcatheter heart valve (THV), with pre- and post-TAVI computed tomography (CT)
- Participants
- 175 patients
- Exposure
- Inflow level vs leaflet level of the Evolut THV
- Outcome
- Stent frame expansion and ellipticity
Methodological detailspre- and post-TAVI CT
- Pre- and post-TAVI CT
Main finding
As reported in the abstract
| Group | Expansion | Ellipticity |
|---|---|---|
| Inflow level | 79.9% ± 6.7% | 1.3 ± 0.3 |
| Leaflet level | 95.4% ± 3.1% | 1.2 ± 0.1 |
Expansion
Inflow level79.9% ± 6.7%
Leaflet level95.4% ± 3.1%
Ellipticity
Inflow level1.3 ± 0.3
Leaflet level1.2 ± 0.1
- Inflow level vs Leaflet level: P < 0.001 — greater under-expansion at inflow
- Inflow level vs Leaflet level: P < 0.001 — greater ellipticity at inflow
Observational data: associations, not effects.
Further findings
Leaflet level expansion and ellipticity by inflow expansion/ellipticity
- Leaflet level ellipticity increased with inflow under-expansion <75%
- P = 0.002
- Leaflet level ellipticity increased with inflow ellipticity ≥1.3
- P < 0.001
- Leaflet level expansion unaffected by inflow expansion or ellipticity
Valve haemodynamics (mean transprosthetic gradient, paravalvular leak) and hypoattenuated leaflet thickening · short-term follow-up
- Neither inflow under-expansion nor ellipticity impacted haemodynamic valve performance or leaflet thickening
Original publication
Stent frame deformation of self-expanding transcatheter heart valves in bicuspid aortic stenosis and impact on valve performance. Eur Heart J Cardiovasc Imaging. Published 3 July 2025. doi:10. 1093/ ehjci/ jeaf195
Composed by AI from the published abstract and checked against it. How Briefs are made →