Arrhythmias & electrophysiology

Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement

EuropacePublished Retrospective cohort study

Authors' conclusion

This study shows a high DFT conversion rate at elective S-ICD generator replacement, which is comparable to conversion rates at implant, despite a rise in shock impedance.

Study context

Population
Patients undergoing elective subcutaneous implantable cardioverter-defibrillator (S-ICD) generator replacement for battery depletion with defibrillation test (DFT), February 2015–June 2022
Participants
121 patients
Exposure
DFT at S-ICD generator replacement vs at implant
Outcome
DFT conversion efficacy (unsuccessful when two conversions at 65 J failed)
Methodological detailssingle-centre (tertiary centre)
  • Single-centre (tertiary centre)

Main finding

As reported in the abstract

DFT conversion success at elective S-ICD generator replacement

DFT successful after first test

95%

DFT successful after two consecutive tests

98%

Observational data: associations, not effects.

Further findings

Shock impedance · significant rise

P < 0.001

Unsuccessful DFT at 65 J

Both patients with unsuccessful DFT at 65 J converted with 80 J
2

Original publication

Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement. Europace. Published 28 June 2023. doi:10.1093/europace/euad184

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.