Arrhythmias & electrophysiology

Thermal side effects during pulsed field ablation: an analysis using computer modelling

EuropacePublished Computer modelling study

Authors' conclusion

Thermal effects from PFA depend on the total energy deposited, with peak voltage being a significant factor. Current commercially available PFA systems have the potential to induce collateral thermal injury with repeated applications of pulsed field energy.

Study context

Population
Simulated myocardial and oesophageal tissue (left atrium model)
Comparison
Pulsed field ablation (PFA) at peak voltages of 1, 1.5 and 2 kV
Outcome
Thermal effects of PFA on myocardial and oesophageal tissue
Follow-up
After a single pulse train
Methodological detailsthree-dimensional computer model of the left atrium, bipolar configuration, monophasic waveform, 100 μs pulse width, 1 s gap, 50 pulses
  • Three-dimensional computer model of the left atrium
  • Bipolar configuration
  • Monophasic waveform, 100 μs pulse width, 1 s gap, 50 pulses

Main finding

As reported in the abstract

Oesophageal temperature
GroupValueTemperature rise
1 kVMinimal
1.5 kV46.3 °C
2 kV>62 °C

Value

1 kV— not stated in the abstract
1.5 kV46.3 °C
2 kV>62 °C

Temperature rise

1 kVMinimal
1.5 kV— not stated in the abstract
2 kV— not stated in the abstract

Original publication

Thermal side effects during pulsed field ablation: an analysis using computer modelling. Europace. Published 17 February 2025. doi:10.1093/europace/euaf035

Composed by AI from the published abstract and checked against it. How Briefs are made →