Arrhythmias & electrophysiologyCohort study

Two-year outcomes of Micra AV leadless pacemakers in the Micra AV CED study

EuropacePublished

  • Retrospective cohort study
  • propensity score overlap weight-adjusted
  • Medicare administrative claims data

Authors' conclusion

Patients implanted with Micra AV had lower complications and re-intervention rates at 2 years than patients implanted with a traditional DC-TV pacemaker. All-cause mortality remained higher in Micra AV patients, likely due to their higher comorbidity burden and other differences in baseline characteristics.

Study context

Population
US Medicare patients implanted with a pacemaker in 2020 and 2021
Participants
7,552 vs 110,558
Exposure
Micra AV leadless pacemaker vs dual-chamber transvenous pacemaker
  • Micra AVn = 7,552
  • Dual-chamber transvenous (DC-TV)n = 110,558reference
Outcome
Chronic complications, re-interventions and all-cause mortality
Follow-up
2 years

Main finding

As reported in the abstract

2 years post-implant

Micra AV

5.3%

Complications (adjusted) · Re-intervention (adjusted) 3.5%

DC-TV

9.6%

Complications (adjusted) · Re-intervention (adjusted) 5.6%

Micra AV vs DC-TVHazard ratio (95% CI)P value
complications
0.54(0.49–0.61)
P < 0.0001
re-intervention
0.62(0.54–0.72)
P < 0.0001
all-cause mortality, unadjusted
2.48(2.35–2.62)
P < 0.0001
all-cause mortality, adjusted
1.53(1.44–1.62)
P < 0.0001

Observational data: associations, not effects.

Further findings

Upgrade to cardiac resynchronization therapy · 2 years · adjusted

P = 0.40

Micra AV upgrade to dual-chamber system · 2 years
1.4%

Original publication

Two-year outcomes of Micra AV leadless pacemakers in the Micra AV CED study. Europace. Published 1 November 2024. doi:10.1093/europace/euae273

Summarised from the published abstract; every number is checked against it