Arrhythmias & electrophysiologyCohort study

Upgrading expandable leadless pacemakers from single- to dual-chamber

EuropacePublished

  • Prospective single-group study (upgrade analysis)
  • multicentre
  • Followed 12 months from attempted atrial leadless pacemaker (LP) implantation

Authors' conclusion

These data confirm the feasibility, safety, and long-term effectiveness of adding an atrial LP in patients with a pre-existing ventricular LP to achieve dual-chamber leadless pacing.

Study context

Population
Patients with a prior ventricular leadless pacemaker indicated for a dual-chamber pacemakerattempted upgrade
Participants
35 patients
Outcome
Upgrade success (single- to dual-chamber leadless pacing)

Main finding

As reported in the abstract

Successful upgrade

Upgrade

91.4%

Observational data: associations, not effects.

Further findings

Atrial device electrical measurements · 12 months

Capture threshold at 0.4 ms (mean ± SD)
0.8 ± 0.6 V
Sense amplitude (mean ± SD)
4.5 ± 2.7 mV
  • stabilized within 1 month and remained stable through 12 months

Total estimated battery longevity · evaluated at 12 months

Ventricular LP
11.3 ± 4.2 years
Atrial LP
5.7 ± 2.6 years
  • accounting for time since implant

Safety

Complications (all procedure-related) · 12 months

Patients with complications
5 (14.3%)
Complications
7
Device or procedure-related deaths
0
Atrial LP revision (retrieved and replaced) 10 months after implantation
1

Original publication

Upgrading expandable leadless pacemakers from single- to dual-chamber. Europace. Published 24 February 2026. doi:10.1093/europace/euag023

Summarised from the published abstract; every number is checked against it