Arrhythmias & electrophysiology

3-Year Outcomes From the Amplatzer Amulet Left Atrial Appendage Occluder Randomized Controlled Trial (Amulet IDE)

JACC Cardiovasc IntervPublished Randomised controlled trial

Authors' conclusion

The Amulet occluder demonstrated continued safety and effectiveness with over 96% free of oral anticoagulation usage through 3 years in a high-risk population compared to the Watchman device.

Study context

Population
High-risk patients undergoing left atrial appendage occlusion
Participants
1,878 patients · 108 sites
Comparison
Amulet vs Watchman 2.5 left atrial appendage occluder
Methodological detailsmulticentre, 108 sites
  • Multicentre
  • 108 sites

Primary result

As reported in the abstract

Ischemic stroke or systemic embolism · 3 years

P = 0.69

Further findings

Free of OAC · 3 years

P < 0.01

All strokes, systemic embolism or CV death · 3 years

P = 0.31

All-cause death · 3 years

P = 0.08

CV death · 3 years

P = 0.14

Device factors (device-related thrombus or peridevice leak ≥3 mm) preceding ischemic stroke and cardiovascular (CV) death · 3 years

More frequent with Watchman than Amulet

Safety

Major bleeding · 3 years

P = 0.46

Original publication

3-Year Outcomes From the Amplatzer Amulet Left Atrial Appendage Occluder Randomized Controlled Trial (Amulet IDE). JACC Cardiovasc Interv. Published 14 August 2023. doi:10.1016/j.jcin.2023.06.022

Summarised from the published abstract; every number is checked against it