A direct wire pacing device for transcatheter heart valve and coronary interventions: a first-in-human, multicentre study of the Electroducer Sleeve

EuroInterventionPublished Prospective single-arm pilot study

Authors' conclusion

This first-in-human trial using the Electroducer Sleeve indicated that this novel, purpose-built, temporary pacing device was safe and effective.

Study context

Population
Patients undergoing transcatheter aortic valve implantation (TAVI) or percutaneous coronary interventions (PCI) requiring cardiac pacing39 TAVI, 21 PCI
Participants
60 patients
Outcome
Haematomas or bleeding complications at the device vascular access site
Methodological detailsfirst-in-human, multicentre, non-randomised, 4 centres in France
  • First-in-human
  • Multicentre
  • Non-randomised
  • 4 centres in France

Main finding

As reported in the abstract

Haematomas or bleeding complications at the device access site

EASY grade I/BARC type I haematomas at device access site

2 patients

3.3%

Secondary outcomes

Effectiveness: haemodynamic effect after each spike delivery

Haemodynamic effect after each spike delivery
54 patients (90%)

Safety

Asymptomatic radial artery occlusion

Asymptomatic radial artery occlusion
2 patients (6.3%)

Allergies

No allergies reported

Original publication

A direct wire pacing device for transcatheter heart valve and coronary interventions: a first-in-human, multicentre study of the Electroducer Sleeve. EuroIntervention. Published 20 February 2023. doi:10.4244/eij-d-22-00662

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.