A direct wire pacing device for transcatheter heart valve and coronary interventions: a first-in-human, multicentre study of the Electroducer Sleeve
Authors' conclusion
This first-in-human trial using the Electroducer Sleeve indicated that this novel, purpose-built, temporary pacing device was safe and effective.
Study context
- Population
- Patients undergoing transcatheter aortic valve implantation (TAVI) or percutaneous coronary interventions (PCI) requiring cardiac pacing39 TAVI, 21 PCI
- Participants
- 60 patients
- Outcome
- Haematomas or bleeding complications at the device vascular access site
Methodological detailsfirst-in-human, multicentre, non-randomised, 4 centres in France
- First-in-human
- Multicentre
- Non-randomised
- 4 centres in France
Main finding
As reported in the abstract
Haematomas or bleeding complications at the device access site
EASY grade I/BARC type I haematomas at device access site
2 patients
3.3%
Secondary outcomes
Effectiveness: haemodynamic effect after each spike delivery
- Haemodynamic effect after each spike delivery
- 54 patients (90%)
Safety
Asymptomatic radial artery occlusion
- Asymptomatic radial artery occlusion
- 2 patients (6.3%)
Allergies
- No allergies reported
Original publication
A direct wire pacing device for transcatheter heart valve and coronary interventions: a first-in-human, multicentre study of the Electroducer Sleeve. EuroIntervention. Published 20 February 2023. doi:10. 4244/ eij-d-22-00662