Accurate detection of lead malfunction from electrocardiogram (ECG)-derived bipolar pacing stimulus amplitude

Heart RhythmPublished Diagnostic accuracy study

Authors' conclusion

The maximum amplitude of a bipolar pacing stimulus on the ECG is significantly lower in normal functioning leads compared with those with known malfunction. This simply derived variable demonstrated good accuracy at identifying lead failure due to insulation breach.

Study context

Population
Transvenous pacing leads, with known malfunction or normal
Participants
138 transvenous leads
Comparison
ECG bipolar pacing stimulus amplitude in malfunctioning vs normal transvenous leads
  • Malfunctionn = 88
  • Normaln = 50reference
Outcome
ECG bipolar stimulus amplitude and diagnostic accuracy for lead malfunction
Follow-up
At time of identification of failure

Performance

As reported in the abstract

at time of identification of failure

P < 0.001

AUC, absolute ECG amplitude
0.93 (95% CI 0.891-0.969)
Sensitivity, cutoff 5 mm/V normalized for output
91%
Specificity, cutoff 5 mm/V normalized for output
92%
AUC, normalized for output
0.967 (95% CI 0.938-0.996)

Original publication

Accurate detection of lead malfunction from ECG-derived bipolar pacing stimulus amplitude. Heart Rhythm. Published 2 April 2024. doi:10.1016/j.hrthm.2024.03.1814