Coronary disease & ACS

Adenosine Contrast Correlations in Evaluating Revascularization: The (ACCELERATION) Study

Circ Cardiovasc IntervPublished investigator-initiated, multicenter, prospective, single-arm trial

Authors' conclusion

cFFR utilizing standardized parameters for contrast delivery leads to clinically acceptable levels of diagnostic performance compared with traditional aFFR to identify physiologically significant intermediate lesions. Future data evaluating the impact on clinical outcomes of cFFR-guided percutaneous coronary intervention are warranted.

Study context

Population
Patients with stable coronary artery disease and intermediate stenosis (40% to 70%)patients, 192 lesions included in the per-protocol analysis · 69 (adenosine fractional flow reserve (aFFR) ≤0.80) · 109 (aFFR >0.80)
Participants
178 patients
Exposure
Diagnostic performance of contrast fractional flow reserve (cFFR) using an automated contrast injector with a standardized volume and rate of delivery of contrast to the reference standard aFFR
Outcome
Diagnostic performance of cFFR, cutoff ≤0.83
Methodological details5 sites across the United States, per-protocol analysis
  • 5 sites across the United States
  • Per-protocol analysis

Main finding

As reported in the abstract

Diagnostic performance of cFFR, cutoff ≤0.83
GroupAccuracySensitivitySpecificity
Contrast fractional flow reservecFFR0.89 (95% CI, 0.83-0.93)0.70 (95% CI, 0.58-0.81)0.99 (95% CI, 0.95-1.00)

Contrast fractional flow reservecFFR

Accuracy0.89 (95% CI, 0.83-0.93)
Sensitivity0.70 (95% CI, 0.58-0.81)
Specificity0.99 (95% CI, 0.95-1.00)

Observational data: associations, not effects.

Further findings

OutcomecFFRMean difference (CI)
Mean difference between cFFR and aFFRBland-Altman analysis0.05 (−0.04 to 0.13)
Diagnostic performance of cFFR, threshold ≤0.85 (highest accuracy in predicting aFFR ≤0.80)
Accuracy0.90 (95% CI, 0.84-0.94)
Sensitivity0.87 (95% CI, 0.77-0.94)
Specificity0.91 (95% CI, 0.84-0.95)
  • Mean difference between cFFR and aFFR: difference of cFFR minus aFFR, as printed "between cFFR and aFFR"; no sign printed; interval level not stated

Original publication

Adenosine Contrast Correlations in Evaluating Revascularization: The (ACCELERATION) Study. Circ Cardiovasc Interv. Published 24 April 2025. doi:10.1161/circinterventions.125.015240

Composed by AI from the published abstract and checked against it. How Briefs are made →