Pulmonary vascular disease & thromboembolism

Agreement of thermodilution and direct Fick methods for cardiac output across varying haemodynamic conditions

ESC Heart FailPublished Retrospective analysis of right heart catheterizations

Authors' conclusion

Although TD shows a predictable bias across flow states, it remains a clinically reliable method for estimating elevated PVR. These findings support the clinical utility of TD in PH assessment, while emphasizing the need for direct Fick measurement in low-output states-particularly in patients evaluated for advanced therapies such as transplantation.

Study context

Population
Patients undergoing right heart catheterization (2014–22)
Participants
852 right heart catheterizations
Index test or model
QTD with Fick-derived Q (QF) across haemodynamic states; PVRTD against elevated PVRF
Target condition
Elevated PVRF at both 2 and 5 Wood unit thresholds
Methodological detailssimultaneous QTD and QF measurements
  • Simultaneous QTD and QF measurements

Performance

As reported in the abstract

Area under the curve
0.97
P
P < .001

Further findings

Correlation of QTD with QF

R
0.78
P
P < 0.001

Bias of QTD vs QF by flow state

L/min

Low-flow states (overestimating Q)
+0.1
Normal/high flow (underestimating Q)
−0.48
  • Flow-dependent bias

Agreement between QTD and QF by tricuspid regurgitation severity

Unaffected by tricuspid regurgitation severity

Patient characteristics

Heart failure
64%
Pulmonary arterial hypertension
20%

Original publication

Agreement of thermodilution and direct Fick methods for cardiac output across varying haemodynamic conditions. ESC Heart Fail. Published 8 January 2026. doi:10.1093/eschf/xvaf001

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.