Arrhythmias & electrophysiology

Antitachycardia pacing success in implantable cardioverter-defibrillators by patient, device, and programming characteristics

Heart RhythmPublished Secondary analysis of the PainFree SmartShock Technology study

Authors' conclusion

Delivering more ATP sequences resulted in a higher overall success of terminating VTs, while programming more ATP was associated with decreased shock burden and no evidence of increased syncope or acceleration.

Study context

Population
Patients with implantable cardioverter-defibrillator (ICD) or CRT-D; spontaneous antitachycardia pacing (ATP)-treated monomorphic VT episodes
Participants
2,770 patients
Comparison
ATP success by patient, device and programming characteristics
  • ICDn = 1,699
  • CRT-Dn = 1,071
Primary endpoint
ATP success rate for terminating VT
Methodological detailsretrospective
  • Retrospective

Primary result

As reported in the abstract

ATP success

P = 0.81

ATP success for terminating ventricular tachycardias (VTs) overall
>80%
Cumulative ATP success at 1 sequence
71%
Cumulative ATP success at ≥8 sequences
87%
ATP success, median cycle length ≥320 ms (significantly higher)
88.0% (95% CI 84.8%-90.6%)

Further findings

ATP success, primary vs secondary prevention (ICD)

P = 0.25

Shock burden

More ATP sequences programmed associated with lower shock burden
P = 0.0005

Safety

No evidence more sequences associated with higher syncope
P = 0.16

Original publication

Antitachycardia pacing success in implantable cardioverter-defibrillators by patient, device, and programming characteristics. Heart Rhythm. Published 19 October 2022. doi:10.1016/j.hrthm.2022.10.015

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.