Arrhythmias & electrophysiology
Artificial intelligence for detection of ventricular oversensing: Machine learning approaches for noise detection within nonsustained ventricular tachycardia episodes remotely transmitted by pacemakers and implantable cardioverter-defibrillators
Authors' conclusion
Machine learning algorithms were accurate in detecting nonphysiological signals within electrograms transmitted by PMs and ICDs.
Study context
- Population
- Pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs) (Boston Scientific) from 4 French hospitals with ≥1 remotely transmitted nonsustained ventricular tachycardia (NSVT) episode
- Participants
- 807 devices
- Comparison
- 4 machine learning algorithms vs remote monitoring team analysis
- Outcome
- Classification of NSVT episodes as physiological or nonphysiological, by F2 score (favoring sensitivity over PPV)
Methodological detailsmulticentre, independent test set
- Multicentre
- Independent test set
Performance
As reported in the abstract
F2 score for nonphysiological signals
| Test or model | F2 score | Sensitivity | Specificity |
|---|---|---|---|
| Remote monitoring team | 0.932 | 95% | 99% |
| 2DTF-CNN | 0.914 | ||
| 2D-DenseNet | 0.906 | ||
| 2DTF-VGG | 0.863 | ||
| 1D-AgResNet | 0.791 |
F2 score
Remote monitoring team0.932
2DTF-CNN0.914
2D-DenseNet0.906
2DTF-VGG0.863
1D-AgResNet0.791
Sensitivity
Remote monitoring team95%
2DTF-CNNNot reported in the abstract
2D-DenseNetNot reported in the abstract
2DTF-VGGNot reported in the abstract
1D-AgResNetNot reported in the abstract
Specificity
Remote monitoring team99%
2DTF-CNNNot reported in the abstract
2D-DenseNetNot reported in the abstract
2DTF-VGGNot reported in the abstract
1D-AgResNetNot reported in the abstract
Further findings
Devices transmitting nonphysiological signals
- Devices with nonphysiological NSVT recordings
- 87 (10.8%)
- NSVT recordings with nonphysiological signals
- 544
Original publication
Artificial intelligence for detection of ventricular oversensing: Machine learning approaches for noise detection within nonsustained ventricular tachycardia episodes remotely transmitted by pacemakers and implantable cardioverter-defibrillators. Heart Rhythm. Published 3 July 2023. doi:10. 1016/ j. hrthm. 2023. 06. 019
Summarised from the published abstract; every number is checked against it