Arrhythmias & electrophysiology

Atrial substrate characterization based on bipolar voltage electrograms acquired with multipolar, focal and mini-electrode catheters

EuropacePublished Mechanistic study

Authors' conclusion

When measuring LA BVA, significant differences were seen between focal, multielectrode, and minielectrode catheters.

Study context

Population
Patients undergoing high-resolution electroanatomic mapping of the left atrial (LA) (72% men, age 68 ± 15 years)
Participants
25 patients
Comparison
Bipolar voltage (BV) amplitude across catheters with different electrode design
Outcome
Equivalent BVA cutoffs per catheter for established reference cutoffs (0.1, 0.2, 0.5, 1.0, 1.5 mV) for LA scar and healthy tissue

Main finding

As reported in the abstract

Equivalent BVA cutoffs vs Nav reference
Scar, Nav 0.5 mV cutoff: Orion equivalent
0.35 mV
Scar, Nav 0.5 mV cutoff: Lasso equivalent
0.48 mV
Scar, 0.2 mV cutoff: Orion equivalent
0.09 mV
Scar, 0.2 mV cutoff: Lasso equivalent
0.14 mV
Healthy tissue, 1.5 mV cutoff: Orion equivalent
1.68 mV
Healthy tissue, 1.5 mV cutoff: Lasso equivalent
2.21 mV

Original publication

Atrial substrate characterization based on bipolar voltage electrograms acquired with multipolar, focal and mini-electrode catheters. Europace. Published 11 May 2023. doi:10.1093/europace/euad127