Coronary disease & ACS

Biolimus-coated versus paclitaxel-coated balloons for coronary in-stent restenosis (BIO ASCEND in-stent restenosis (ISR)): a randomised, non-inferiority trial

EuroInterventionPublished Randomised controlled trial

Primary resultResult — go to the resultMean difference −0.02 mm 95% CI −0.12 to 0.07

Authors' conclusion

This head-to-head comparison of a novel BCB shows similar angiographic outcomes in the treatment of coronary DES-ISR compared with a clinically proven PCB.

Study context

Population
Patients with coronary drug-eluting stent (DES) in-stent restenosis
Participants
280 patients17 centres
Comparison
Biolimus-coated balloon vs paclitaxel-coated balloon
  • Biolimus-coated balloon (BCB)n = 140
  • Paclitaxel-coated balloon (PCB)n = 140reference
Primary endpoint
In-segment late lumen loss (independent core laboratory)
Follow-up
9 months
Methodological detailsprospective, multicentre, non-inferiority
  • Prospective
  • Multicentre
  • Non-inferiority

Primary result

As reported in the abstract

Mean difference

−0.02 mm

95% CI −0.12 to 0.07

P < 0.0001 for non-inferiority

By group

BCBvs PCB
mean difference −0.02 mm (95% CI −0.12 to 0.07); P < 0.0001 for non-inferiority

Further findings

OutcomeBCBPCBMean difference (mm²) (95% CI)P value
Neointimal area (OCT substudy) · 9 monthsPrespecified subgroup2.32 ± 1.04 mm²2.37 ± 0.93 mm²−0.09 (−0.94 to 0.76)P = non-significant

Clinical outcomes · 12 months

Similar clinical outcomes in both groups

Original publication

Biolimus-coated versus paclitaxel-coated balloons for coronary in-stent restenosis (BIO ASCEND ISR): a randomised, non-inferiority trial. EuroIntervention. Published 1 July 2024. doi:10.4244/eij-d-24-00295

Composed by AI from the published abstract and checked against it. How Briefs are made →