Cardiac magnetic resonance imaging at 3T in patients with magnetic resonance imaging-conditional cardiac implantable electronic devices

Heart RhythmPublished Retrospective cohort study

Authors' conclusion

Our single-center real-world experience confirms the safety of CIED-CMR at 3T for appropriately labeled CIEDs, with no adverse events or changes to device parameters requiring reprogramming. Although images may be suboptimal owing to artifacts, more than 90% of scans were diagnostic.

Study context

Population
Patients with MRI-conditional cardiac implantable electrical devices (CIEDs) referred for clinical CMR at 3T, September 2020 to February 2024
Participants
87 patients90 scans
Outcome
Safety and quality of CMR at 3T
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Adverse events during any scan
0
Changes in battery voltage or lead parameters requiring device reprogramming
0
Diagnostic scans
82 (91%)
Good or good/intermediate quality, cine sequences
50 (59%)
Good or good/intermediate quality, late gadolinium enhancement (LGE) sequences
49 (58%)
No artifacts affecting interpretation, cine sequences
73 (86%)
No artifacts affecting interpretation, LGE sequences
47 (55%)

Original publication

Cardiac magnetic resonance imaging at 3T in patients with magnetic resonance imaging-conditional cardiac implantable electronic devices. Heart Rhythm. Published 23 July 2025. doi:10.1016/j.hrthm.2025.07.032

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.