Arrhythmias & electrophysiology
Chronic wireless communication between dual-chamber leadless pacemaker devices
Authors' conclusion
Wireless implant-to-implant communication in the new dual-chamber leadless pacemaker system demonstrated successful transmissions in >90% of beats throughout the 6-month evaluation period. Communication success improved significantly over time postimplant for specific subgroups.
Study context
- Population
- Patients meeting standard dual-chamber pacing indications, de novo dual-chamber leadless pacemaker (LP) implant
- Participants
- 399 patients
- Outcome
- Chronic i2i transmission success rates (A-to-V and V-to-A)
- Follow-up
- Implant, discharge, 1, 3 and 6 months postimplant
Methodological detailsAveir DR i2i Study, international
- Aveir DR i2i Study
- International
Main finding
As reported in the abstract
i2i success rate · implant to 6M
Median ALP-to-VLP (A-to-V) and VLP-to-ALP (V-to-A) i2i success rates exceeded 90% of beats at all time-points
>90%
Observational data: associations, not effects.
Further findings
Patients with i2i success in <70% of beats at implant (subgroup) · implant, 1M, 6M
- A-to-V <70% at implant
- 19%
- V-to-A <70% at implant
- 31%
- Improvement by 1M
- roughly 40%
- Patients <70% by 6M
- roughly 5%
Original publication
Chronic wireless communication between dual-chamber leadless pacemaker devices. Heart Rhythm. Published 19 October 2024. doi:10. 1016/ j. hrthm. 2024. 10. 020
Summarised from the published abstract; every number is checked against it