Authors' conclusion

THV deformation, i.e., underexpansion and an hourglass shape, commonly occurs after MViV implantation and is negatively affected by excessive oversizing.

Study context

Population
Patients undergoing mitral valve-in-valve with SAPIEN prostheses
Participants
33 patients
Outcome
Extent and predictors of THV deformation (deformation index, eccentricity, neosinus volume, asymmetric leaflet expansion, vertical deformation) and HALT on cardiac CT
Follow-up
30 days post-procedure
Methodological detailsComputed tomography (CT) performed prospectively at 30 days
  • CT performed prospectively at 30 days

Main finding

As reported in the abstract

THV deformation · 30 days
Expansion area at narrowest level (median, IQR 68.1-84.1)
74.0%
Transcatheter heart valves (THV) deformation index (median, IQR 1.13-1.29)
1.21
Eccentricity range
0.24 to 0.28
Oversizing associated with greater underexpansion
β=−0.634; p<0.001
Oversizing associated with higher deformation index
β=0.594; p<0.001

Observational data: associations, not effects.

Further findings

Hypoattenuating leaflet thickening (HALT) on 30-day CT

HALT incidence (9 of 33)
27.3%
On anticoagulation
32 of 33
Prothrombin time and international normalised ratio (PT-INR) <2.5 at CT
23 of 32
With PT-INR <2.5, HALT predominantly with THV deformation index ≥1.18

Original publication

Deformation of transcatheter heart valves with mitral valve-in-valve. EuroIntervention. Published 18 December 2023. doi:10.4244/eij-d-23-00614