Arrhythmias & electrophysiologyCohort study

Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation

EuropacePublished

  • Observational study

Authors' conclusion

A rapid and enduring drop of the LI may predict effective lesion formation, while slightly changing or unchanged LI is associated with unsuccessful applications. A moderate LI drop during the first 4 s of radiofrequency application predicts ineffective radiofrequency delivery.

Study context

Population
Point-by-point pulmonary vein isolation with a contact force sensing catheter559 successful, 84 unsuccessful applications
Participants
559 vs 84
Exposure
Successful vs unsuccessful RF applications
  • Successfuln = 559
  • Unsuccessfuln = 84reference
Outcome
Temporal characteristics of LI drop during ablation and correlation with acute lesion efficacy

Main finding

As reported in the abstract

LI drop and lesion efficacy

Successful

161-150-141 Ω

Local impedance (LI) baseline-2 s-10 s

Unsuccessful

153 vs. 145 Ω

LI baseline vs 2 s

Unsuccessful vs UnsuccessfulOdds ratio (CI)P value
failing to reach the <9 Ω cut-point predicts unsuccessful application
3.82(2.34–6.25)
P < 0.001
LI drop <9 Ω at 4 s for unsuccessful application
AUC 0.73 (CI 0.67–0.76); P < 0.001
Successful: higher baseline LI
P < 0.001
Successful: larger LI drop during ablation
P < 0.001
Unsuccessful: drop baseline to 2 s
P < 0.001
Unsuccessful: no further LI change after 2 s
P = 0.99
Successful: further significant LI decrease
P < 0.001

Observational data: associations, not effects.

Original publication

Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation. Europace. Published 7 October 2024. doi:10.1093/europace/euae260