Arrhythmias & electrophysiology

Feasibility and safety of single-pulse ablation in 20 patients with atrial fibrillation

Heart RhythmPublished Prospective single-group feasibility and safety study

Authors' conclusion

Acute electrical PVI could be achieved safely and rapidly. Acute silent cerebral lesions were detected in 3 of 20 patients (15%) and may be caused by ablation or changes of therapeutic and diagnostic catheters over a single transseptal access.

Study context

Population
Patients with symptomatic AF undergoing pulmonary vein isolation
Participants
20 patients
Outcome
Safety of single-pulse ablation for PVI

Main finding

As reported in the abstract

All pulmonary veins successfully isolated

20 patients

Irreversible electroporation (IRE) applications per patient

11.8 ± 1.4 (mean)

PV reconnection during adenosine testing; reisolated with 2 additional IRE pulses

1

Periprocedural complications

0

Observational data: associations, not effects.

Safety

Silent cerebral lesions (brain magnetic resonance imaging (MRI)) · day 1 after ablation

Punctate asymptomatic lesions, 3 of 20 patients
15%
  • At follow-up MRI, the lesion disappeared in 1 patient; 1 lesion persisted in the other 2 patients

Asymptomatic esophageal lesion (esophagoscopy) · day 1

Asymptomatic esophageal lesion, 1 of 20 patients
5%
  • Resolved completely at repeat esophagoscopy on day 22

Periprocedural complications

No periprocedural complications observed

Original publication

Feasibility and safety of single-pulse ablation in 20 patients with atrial fibrillation. Heart Rhythm. Published 28 May 2025. doi:10.1016/j.hrthm.2025.05.054

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.