Valvular & structural heart diseaseResearch article

Fibrotic Tissue Ingrowth of Transcatheter Heart Valve Frames

JACC Cardiovasc IntervPublished

  • Explant pathology study

Authors' conclusion

Fibrotic THV frame ingrowth is a common phenomenon that may complicate THV explantation and impede neosinus washout and coronary access after TAVR.

Study context

Population
Explanted transcatheter heart valves after transcatheter aortic valve replacement (TAVR)explanted transcatheter heart valve (THVs)
Participants
34 vs 10
Comparison
SAPIEN vs Evolut/CoreValve explanted THVs
  • SAPIENn = 34
  • EV/CVn = 10
Outcome
THV frame obstruction due to fibrotic tissue ingrowth and its potential impact on coronary access and neosinus flow

Main finding

As reported in the abstract

Frame fibrotic ingrowth
Frame with fibrotic ingrowth, all explanted THVs
79.5%
Implant duration correlated with obstruction severity
P = 0.003
Cells in likely cannulation areas not passable by 6-F catheter
13.0%
Redo-TAVR models unable to be fenestrated by 6-F catheter
up to 76%
Median granulation tissue thickness
0.5 mm

Further findings

Distribution and clinical correlates

Ingrowth more common in upper and lower regions of EV/CVs and upper cells adjacent to commissures in SAPIEN
Greater ingrowth associated with damage to native structure at explantation
Frame adhesions not detectable by standard clinical imaging

Original publication

Fibrotic Tissue Ingrowth of Transcatheter Heart Valve Frames. JACC Cardiovasc Interv. Published 23 February 2026. doi:10.1016/j.jcin.2025.11.003

Summarised from the published abstract; every number is checked against it