Redo-TAVI with the SAPIEN 3 valve in degenerated calcified CoreValve/Evolut explants

EuroInterventionPublished Mechanistic study

Authors' conclusion

S3 implantation at node 5 in calcified CV/EV valves resulted in satisfactory hydrodynamic performance in most configurations tested with stable leaflet overhang throughout the cardiac cycle.

Study context

Population
Explanted degenerated calcified CoreValve/Evolut transcatheter aortic valves (TAVs) with SAPIEN 3 implanted at node 5 (redo-TAVI)S3 TAVs in CV/EV explants
Participants
4 participants
Outcome
S3 hydrodynamic performance (ISO 5840-3) in calcified CV/EV explants
Methodological detailsex vivo hydrodynamic testing, ISO 5840-3 standard
  • Ex vivo hydrodynamic testing
  • ISO 5840-3 standard

Main finding

As reported in the abstract

All S3 implants showed adequate effective orifice area (EOA)
4
Regurgitant fraction (RF) within accepted value (<20%)
3 out of 4
CV/EV leaflet overhang
25-37%
Calcified leaflets stationary through cardiac cycle
difference <9%
Leaflet pinwheeling from S3 frame underexpansion
13-30%

Original publication

Redo-TAVI with the SAPIEN 3 valve in degenerated calcified CoreValve/Evolut explants. EuroIntervention. Published 18 November 2024. doi:10.4244/eij-d-24-00619