Arrhythmias & electrophysiology

Frequency Domain Analysis of Endocardial Electrograms for Detection of Nontransmural Myocardial Fibrosis in Nonischemic Cardiomyopathy

JACC Clin ElectrophysiolPublished Diagnostic accuracy study

Authors' conclusion

These data indicate the promising discriminatory value of endocardial EGM frequency content in the assessment of concealed myocardial substrate.

Study context

Population
Nonischemic cardiomyopathy patients undergoing VT ablation with preprocedural cardiac CT and late iodine enhancement
Participants
13 patients6,015 EGM pairs
Comparison
Frequency-domain EGM features vs voltage amplitude for predicting MMF
Outcome
Prediction of midmyocardial fibrosis (MMF) presence (AUC)
Methodological detailssupervised machine learning, stratified k-fold cross-validation within patients
  • Supervised machine learning
  • Stratified k-fold cross-validation within patients

Performance

As reported in the abstract

Mean AUC for MMF prediction
Test or modelMean AUC
Frequency features0.841 (95% CI: 0.789-0.884)
Voltage0.591 (95% CI: 0.530-0.658)

Mean AUC

Frequency features0.841 (95% CI: 0.789-0.884)
Voltage0.591 (95% CI: 0.530-0.658)
  • Frequency features vs Voltage: P < 0.001 — frequency-trained classifiers better predicted MMF

Original publication

Frequency Domain Analysis of Endocardial Electrograms for Detection of Nontransmural Myocardial Fibrosis in Nonischemic Cardiomyopathy. JACC Clin Electrophysiol. Published 18 January 2023. doi:10.1016/j.jacep.2022.11.019

Composed by AI from the published abstract and checked against it. How Briefs are made →