Heart failure & cardiomyopathies

Heart sound–derived systolic time intervals for atrioventricular delay optimization in cardiac resynchronization therapy

Heart RhythmPublished Mechanistic study

Authors' conclusion

AV delays advised by PCG-derived STIs cause only a minimal loss of pump function compared with those based on invasive hemodynamic measurements.

Study context

Population
Pigs with atrioventricular (AV) block undergoing biventricular pacing (BiVP) and patients undergoing CRT device implantation5 pigs · 21 patients
Comparison
Phonocardiography (PCG)-derived STI-guided AV delay vs invasive haemodynamic-guided AV delay during BiVP
Outcome
Error of STI-predicted vs haemodynamically optimal AV delay
Methodological detailsanimal and patient study
  • Animal and patient study

Main finding

As reported in the abstract

Error in predicting optimal AV delay
Pigs: VS1 and S1S2 median error vs optimal AV delay
2-28 ms
Pigs: median loss of pump function
<2%
Patients: VS1 error
32.5 ms
Patients: S1S2 error
37.5 ms
Patients: lower left ventricular pressure (LVP) and stroke work
0.2%-0.9%

Further findings

Errors in patients with full-capture AV delay >180 ms

ms

VS1 error
21
S1S2 error
24
  • 8 patients

Original publication

Heart sound–derived systolic time intervals for atrioventricular delay optimization in cardiac resynchronization therapy. Heart Rhythm. Published 24 December 2022. doi:10.1016/j.hrthm.2022.12.031

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.