Arrhythmias & electrophysiology

High predicted durability for the novel small-diameter OmniaSecure defibrillation lead

Heart RhythmPublished Reliability modeling study

Authors' conclusion

Consistent with early clinical trial experience, modeling projects highly durable 10-year performance of the OmniaSecure lead, including within the active adolescent pediatric population, which may uniquely benefit from a novel 4.7F defibrillation lead designed for reliability.

Study context

Population
Patients with ICDs (LEADR trial subset imaged in vivo)
Outcome
Projected fracture-free survival
Follow-up
10 years
Methodological detailsvalidated reliability model, patient and bench test data, biplane fluoroscopy in LEADR trial subset
  • Validated reliability model
  • Patient and bench test data
  • Biplane fluoroscopy in LEADR trial subset

Main finding

As reported in the abstract

OmniaSecure

98.2%

Further findings

OutcomeValues
Projected fracture-free survival in adolescents · 10 yearsOmniaSecure 97.9%

Original publication

High predicted durability for the novel small-diameter OmniaSecure defibrillation lead. Heart Rhythm. Published 26 September 2024. doi:10.1016/j.hrthm.2024.09.005

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.