Arrhythmias & electrophysiology

Performance and Safety of the Extravascular Implantable Cardioverter Defibrillator Through Long-Term Follow-Up: Final Results From the Pivotal Study

CirculationPublished Prospective single-arm premarket clinical study

Authors' conclusion

From implantation to study completion, the EV ICD Pivotal study demonstrated that a single integrated system with an extravascular lead placed in the substernal space maintains high ATP success, effective defibrillation, and a consistent safety profile.

Study context

Population
Individuals with a Class I or IIa indication for a single-chamber ICD per guidelines
Participants
316 individuals299 successful implantations
Outcome
Freedom from major system- or procedure-related complications; appropriate and inappropriate therapy rates
Follow-up
Through 3 years
Methodological detailsglobal, Kaplan-Meier
  • Global
  • Kaplan-Meier

Main finding

As reported in the abstract

Freedom from system- or procedure-related major complications at 1 year
91.9%
Freedom from system- or procedure-related major complications at 3 years
89.0%
First any appropriate therapy at 3 years (KM-estimated)
9.2%
Inappropriate shock rate at 1 year
9.8%
Inappropriate shock rate at 3 years (P-wave oversensing predominant cause)
17.5%

Observational data: associations, not effects.

Further findings

Anti-tachycardia pacing (ATP) success

ATP successful in episodes (37/48)
77.1%
ATP use significantly increased from discharge to last follow-up
P < 0.0001

Shock success in discrete, spontaneous ventricular arrhythmias

Shock therapy successful (27/27)
100%

Spontaneous arrhythmic episodes appropriately treated

Patients with appropriately treated episodes
24
Spontaneous arrhythmic episodes
82
Treated with ATP only (38 episodes)
46.3%
Treated with shock only (34 episodes)
41.5%
Treated with both (10 episodes)
12.2%
  • 299 patients with successful implantation

Safety

Major intraprocedural complications
0
Lead dislodgement (10 events; n=9 patients)
2.8%
Postoperative wound or device pocket infection (n=8)
2.5%
Device inappropriate shock delivery (n=4)
1.3%
System revisions due to major complications
24

Original publication

Performance and Safety of the Extravascular Implantable Cardioverter Defibrillator Through Long-Term Follow-Up: Final Results From the Pivotal Study. Circulation. Published 26 September 2024. doi:10.1161/circulationaha.124.071795

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.