Impact of echocardiographic analyses of valvular event timing on myocardial work indices

Eur Heart J Cardiovasc ImagingPublished Methodological study

Authors' conclusion

Myocardial work indices differ depending on the event timing modality used, with visual assessment yielding lower GWI values compared with Doppler-based methods.

Study context

Population
Healthy subjects and patients with aortic valve sclerosis, AF or reduced LVEF (50 each); median age 72 years, 50% men50 per subgroup
Participants
200 subjects
Comparison
Valvular event timing by spectral Doppler vs colour tissue Doppler imaging (TDI) M-mode vs visual assessment for pressure-strain loop (PSL) myocardial work
Outcome
Myocardial work indices, incl. global work index (GWI)

Main finding

As reported in the abstract

GWI

Mean difference

114 mmHg%

P < 0.001

Spectral Doppler vs Visual assessment · ± 93

Mean difference

83 mmHg%

P < 0.001

TDI vs Visual assessment · ± 90

Mean difference

30 mmHg%

P < 0.001

Spectral Doppler vs TDI · ± 53

Further findings

Global work index (GWI), healthy subgroup

OutcomeMean difference (mmHg%)P value
Spectral Doppler vs Visual assessment · ± 77 · systematic bias160P < 0.001
TDI vs Visual assessment · ± 74124P < 0.001

Original publication

Impact of echocardiographic analyses of valvular event timing on myocardial work indices. Eur Heart J Cardiovasc Imaging. Published 19 August 2022. doi:10.1093/ehjci/jeac171

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.