Arrhythmias & electrophysiologyCohort study

Leadless pacemaker implantation via the internal jugular vein

EuropacePublished

  • Retrospective cohort study
  • two centres
  • consecutive patients

Authors' conclusion

The jugular approach may represent a safe and efficient alternative to femoral implantation of the Micra leadless pacemaker.

Study context

Population
Patients undergoing Micra leadless pacemaker implantation
Participants
100 vs 100
Exposure
Jugular vs femoral leadless pacemaker implantation
  • Jugularn = 100
  • Femoraln = 100reference
Outcome
Peri-procedural safety and efficacy

Main finding

As reported in the abstract

Procedure and fluoroscopy times
GroupProcedure time, mean ± SDFluoroscopy time, mean ± SD
Femoral48.9 ± 21.0 min7.7 ± 7.8 min
  • Jugular vs Femoral: P < 0.01 — procedure and fluoroscopy times both shorter with jugular

Observational data: associations, not effects.

Further findings

Pacing threshold (at 0.24 ms pulse width) and sensed amplitude

Electrical parameters stable at follow-up in all patients; similar between approaches

Safety

OutcomeJugularFemoral
Complications
Pericardial effusion11
Dislocation1Not reported in the abstract
Femoral artery injuriesNot reported in the abstract2
Major groin haematomasNot reported in the abstract13

Original publication

Leadless pacemaker implantation via the internal jugular vein. Europace. Published 25 July 2024. doi:10.1093/europace/euae199

Summarised from the published abstract; every number is checked against it