A leadless ventricular pacemaker providing atrioventricular synchronous pacing in the real-world setting: 12-Month results from the Micra atrioventricular (AV) post-approval registry

Heart RhythmPublished Prospective observational registry

Main finding · Major complicationsResult — go to the resultHazard ratio 0.42 95% CI 0.28–0.61

Authors' conclusion

The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple comorbidities, with a significantly lower rate of complications and system revisions through 12 months compared to a historical cohort of patients with transvenous dual-chamber pacemakers.

Study context

Population
Real-world patients implanted with Micra AV at 97 centres in 19 countries796 implanted
Participants
801 vs 2,667
Exposure
Micra AV vs Transvenous dual-chamber
  • Micra AVn = 801
  • Transvenous dual-chambern = 2,667reference
Outcome
Major complications and system revisions
Follow-up
12 months
Methodological detailssingle-arm, worldwide, interim analysis, historical control cohort
  • Single-arm
  • Worldwide
  • Interim analysis
  • Historical control cohort

Main finding

Major complications

Hazard ratio

0.42

95% CI 0.28–0.61

P < 0.001

By group

Micra AV vs Transvenous dual-chamber
HR 0.42 (95% CI 0.28–0.61); P < 0.001

As reported in the abstract

Observational data: associations, not effects.

Further findings

OutcomeMicra AVTransvenous dual-chamberHR (95% CI)P value
System revisions · 12 months · interim analysis1.5%5.5%0.25 (0.13–0.47)P < 0.001
  • System revisions: largely driven by absence of lead dislodgments requiring revision

AV synchrony index (patients paced >90%)

Median AV synchrony index in patients paced >90% (IQR 65.2%-86.4%)
79.4%

Successful implantation

Device successfully implanted, 796 of 801 patients
99.4%

Safety

OutcomeMicra AVTransvenous dual-chamberHR (95% CI)P value
Major complications · 12 months · interim analysis3.7%8.8%0.42 (0.28–0.61)P < 0.001

Original publication

A leadless ventricular pacemaker providing atrioventricular synchronous pacing in the real-world setting: 12-Month results from the Micra AV post-approval registry. Heart Rhythm. Published 13 June 2024. doi:10.1016/j.hrthm.2024.06.008

Summarised from the published abstract; every number is checked against it