Arrhythmias & electrophysiology

LOCAlizaTion and clinical corrElation of Left Bundle Branch Pacing lead: Insights from a computed tomographic angiography (LOCATE left bundle branch pacing (LBBP)) study

Heart RhythmPublished Prospective imaging study

Authors' conclusion

Deep septal deployment of the lead in the LV subendocardium for LBB capture is safe, with 94% of patients showing consistent conduction system capture during follow-up.

Study context

Population
Consecutive patients with successful LBBP using a lumenless lead (excluded: no left bundle branch (LBB) capture, contrast allergy, renal dysfunction)of 105 with computed tomographic angiography (CTA) after LBBP
Participants
67 patients
Outcome
Consistent LBB capture with helix tip <2 mm from LV blood pool by CTA
Follow-up
6 months
Methodological detailsconsecutive patients, CT angiography follow-up, Mean follow-up 33.8 ± 4.4 months
  • Consecutive patients
  • CT angiography follow-up
  • Mean follow-up 33.8 ± 4.4 months

Main finding

As reported in the abstract

Primary endpoint achieved

Primary endpoint achieved

89.5%

n = 60

Observational data: associations, not effects.

Secondary outcomes

Loss of conduction system capture (LOCSC) predictors

OR (95% CI)P value
Lack of nonselective to selective capture transition during implantation18.0 (1.7–192.7)P = 0.01
LOCSC predictors — Lead displacement >2 mm away from left ventricular (LV) blood pool: sensitivity
100%
LOCSC predictors — Lead displacement >2 mm away from LV blood pool: specificity
95%

Consistent LBB capture (group I) · 6 months

Consistent LBB capture
94% (n = 63)

Loss of conduction system capture (LOCSC, group 2) · 6 months

LOCSC
6% (n = 4)

Lead stability: mean distance LV blood pool to helix tip, postimplant vs 6 months

No difference in mean distance (−0.5 ± 1.8 mm vs −0.1 ± 2.1 mm)
P = 0.23

Nonselective to selective capture transition

Nonselective to selective capture transition
82% (n = 55)

Safety

Perforation into LV cavity with complete loss of capture · 6 months

No perforation into the LV cavity with complete loss of capture

Original publication

LOCAlizaTion and clinical corrElation of Left Bundle Branch Pacing lead: Insights from a computed tomographic angiography (LOCATE LBBP) study. Heart Rhythm. Published 26 November 2024. doi:10.1016/j.hrthm.2024.11.038

Composed by AI from the published abstract and checked against it. How Briefs are made →