Non-invasive three-dimensional electrical activation mapping to predict cardiac resynchronization therapy response: site of latest left ventricular activation relative to pacing site
Authors' conclusion
The distance between LV pacing site and latest electrical activation is a strong independent predictor for CRT response.
Study context
- Population
- Consecutive cardiac resynchronization therapy (CRT) patients undergoing electrocardiographic Imaging (ECGI) and chest/heart CT 6–24 months post-implant
- Participants
- 111 patients
- Exposure
- Distance of left ventricle (LV) pacing site to latest electrically activated site (ECGI) in CRT responders vs non-responders
- Outcome
- CRT response (LVESV reduction ≥15%)
- Follow-up
- 10 ± 3 months post CRT-implant
Methodological detailsconsecutive patients
- Consecutive patients
Main finding
As reported in the abstract
Distance from LV pacing site (LVPS) to LEAS (dp)
P < 0.001
- Responders
- 67%
- Overall dp
- 42 ± 23 mm
- Longer dp and lack of CRT optimization algorithm use the only independent predictors of non-response (AUC)
- 0.906
- Dp of 47 mm delineated responders and non-responders (AUC)
- 0.931
Observational data: associations, not effects.
Further findings
Latest electrically activated site (LEAS) location
- LEAS outside mid-to-basal lateral segments
- 35%
Original publication
Non-invasive three-dimensional electrical activation mapping to predict cardiac resynchronization therapy response: site of latest left ventricular activation relative to pacing site. Europace. Published 1 March 2023. doi:10. 1093/ europace/ euad041