Valvular & structural heart disease

Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice: one-yearresults of the ACURATE neo2 PMCF Study

EuroInterventionPublished Post-market clinical follow-up study

Authors' conclusion

The study outcomes confirm favourable performance and safety up to 1 year in patients treated with ACURATE neo2 in routine clinical practice.

Study context

Population
Patients with severe AS in routine clinical practice246 implanted
Participants
250 patients
Outcome
Clinical outcomes (VARC-2 safety events) and valve performance
Follow-up
Up to 1 year
Methodological detailssingle-arm, independent core laboratories
  • Single-arm
  • Independent core laboratories

Main finding

As reported in the abstract

VARC-2 clinical outcomes

All-cause mortality at 30 days

0.8%

All-cause mortality at 1 year

5.1%

Stroke at 1 year

3.0%

Disabling stroke at 1 year

1.3%

Further findings

Hypoattenuated leaflet thickening (HALT) >50% involvement of ≥1 leaflet (4D CT)

HALT at 30 days
9%
HALT at 1 year
12%
No association between HALT and 1-year clinical or haemodynamic outcomes

Mean aortic valve gradient · up to 1 year

Baseline (mmHg)
47.6±14.5
1 year (mmHg)
7.6±3.2

Mean aortic valve area · up to 1 year

Baseline (cm²)
0.7±0.2
1 year (cm²)
1.7±0.4

Paravalvular leakage (PVL) · up to 1 year

Mild or no/trace PVL
99%
Moderate PVL
<1%
Severe PVL
0

Original publication

Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice: one-yearresults of the ACURATE neo2 PMCF Study. EuroIntervention. Published 1 January 2024. doi:10.4244/eij-d-23-00823

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.