Arrhythmias & electrophysiologyResearch article

P-wave morphology of Bachmann bundle pacing based on electroanatomic mapping and intracardiac echocardiography

Heart RhythmPublished

  • This study included 20 patients who underwent a catheter ablation for atrial fibrillation

Authors' conclusion

Using a combined definition of the BB using EAM and intracardiac echocardiography, paced P-wave parameters were identified that can be used to confirm BB capture with high sensitivity and specificity, which could be useful when targeting the BB for permanent atrial pacing.

Study context

Population
Patients who underwent a catheter ablation for atrial fibrillation
Participants
20 patients
Comparison
Paced P-wave criteria against the Bachmann bundle (BB) defined anatomically by electroanatomic mapping (EAM) and intracardiac echocardiography

Main finding

As reported in the abstract

Agreement of echocardiographically identified BB with BB determined by EAM
  • In all cases, the echocardiographically identified BB was consistent with the BB determined by EAM

Further findings

Local electrogram (EGMs) at the BBRA

Local EGMs at the BBRA vary morphologically

P-wave morphology of BBRA pacing

P-wave duration shorter than that of the sinus rhythm (SR)
>10 ms
12-lead P-wave axes recapitulated normal sinus rhythm (SR) P waves
Inferior P-wave amplitude increased with BB capture relative to the native SR P wave
No isoelectric interval pacing centrally at the right atrial BB breakout area

Prediction of the BB when all 4 criteria are met

sensitivity
95%
specificity
86%

Original publication

P-wave morphology of Bachmann bundle pacing based on electroanatomic mapping and intracardiac echocardiography. Heart Rhythm. Published 14 May 2026. doi:10.1016/j.hrthm.2026.05.012

Summarised from the published abstract; every number is checked against it