Heart failure & cardiomyopathies

Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation

EuropacePublished Observational cohort study

Authors' conclusion

Patients received more alerts during periods of AF. The ability of the algorithm to identify increased risk of HF events was confirmed during AF, despite a lower IN/OUT-of-alert incidence rate ratio in comparison with non-AF periods and non-AF patients.

Study context

Population
ICD patients with HeartLogic heart failure (HF) monitoring activated, 26 centres
Participants
568 patients · 26 centres
Exposure
HeartLogic performance during atrial high-rate episode (AHRE) burden ≥20 h/day (atrial fibrillation (AF)) vs <1 h/day periods, and vs matched non-AHRE patients
  • AHREn = 53
  • Matched non-AHREn = 106reference
Methodological detailsmulticentre, propensity-score matched (2:1)
  • Multicentre
  • Propensity-score matched (2:1)

Main finding

As reported in the abstract

Alert rate and IN/OUT-of-alert HF hospitalization rates

AHRE

Alert rate, AHRE burden <1 h/day
1.2 (95% CI: 1.0-1.5)/patient-year
Alert rate, AHRE burden ≥20 h/day
2.0 (95% CI: 1.5-2.6)/patient-year
HF hospitalization rate, IN-alert
0.34 (95% CI: 0.15-0.69)/patient-year
HF hospitalization rate, OUT-of-alert
0.06 (95% CI: 0.02-0.14)/patient-year

Matched non-AHRE

HF hospitalization rate, IN-alert
0.29 (95% CI: 0.12-0.60)/patient-year
HF hospitalization rate, OUT-of-alert
0.04 (95% CI: 0.02-0.08)/patient-year

Alert rate, AHRE burden <1 h/day

AHRE1.2 (95% CI: 1.0-1.5)/patient-year
Matched non-AHRENot reported in the abstract

Alert rate, AHRE burden ≥20 h/day

AHRE2.0 (95% CI: 1.5-2.6)/patient-year
Matched non-AHRENot reported in the abstract

HF hospitalization rate, IN-alert

AHRE0.34 (95% CI: 0.15-0.69)/patient-year
Matched non-AHRE0.29 (95% CI: 0.12-0.60)/patient-year

HF hospitalization rate, OUT-of-alert

AHRE0.06 (95% CI: 0.02-0.14)/patient-year
Matched non-AHRE0.04 (95% CI: 0.02-0.08)/patient-year
IN-alert vs OUT-of-alertIncidence rate ratio (95% CI)
AHRE group, AHRE burden <1 h/day
8.59(1.67–55.31)
AHRE group, AHRE burden ≥20 h/day
2.70(1.01–28.33)

Observational data: associations, not effects.

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation. Europace. Published 1 September 2023. doi:10.1093/europace/euad261

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.