Arrhythmias & electrophysiology

The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice

EuropacePublished Retrospective cohort study

Authors' conclusion

In contemporary practice, the PRAETORIAN score can be simplified.

Study context

Population
Patients who underwent subcutaneous implantable cardioverter defibrillator (S-ICD) implantation with intermuscular generator positioning
Participants
1,253 patients · 33 centres
Outcome
Correlation of PRAETORIAN score with clinical characteristics and implantation variables
Methodological detailsmulticentre, 33 centres
  • Multicentre
  • 33 centres

Main finding

As reported in the abstract

PRAETORIAN score

PRAETORIAN score <90 (low risk of conversion failure)

95.7%

Optimal posterior device placement with intermuscular technique

all patients

Significant sub-generator fat

3%

Shock impedance >88 Ohm detecting score ≥90: NPV

98%

95% CI 97–99%

Observational data: associations, not effects.

Original publication

The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice. Europace. Published 21 January 2025. doi:10.1093/europace/euaf011

Summarised from the published abstract; every number is checked against it