Personalized voltage maps guided by cardiac magnetic resonance in the era of high-density mapping

Heart RhythmPublished Diagnostic accuracy study

Authors' conclusion

Personalized voltage CMR-guided personalized voltage maps enable a better identification of the substrate with a higher correlation with both DZs and VT isthmuses than do conventional voltage maps using fixed thresholds.

Study context

Population
Patients with scar-related ventricular tachycardia (VT) undergoing ablation after cardiac magnetic resonance (CMR) (October 2018 to December 2020)25 (78.1%) ischemic cardiomyopathy
Participants
32 patients
Comparison
Personalized CMR-guided voltage thresholds vs standard thresholds
Outcome
Diagnostic yield: DZs identified as border zone tissue and VT isthmuses correlated with a voltage channel
Methodological detailsconsecutive patients
  • Consecutive patients

Performance

As reported in the abstract

DZs identified as border zone tissue (of 52)

VT isthmuses correlated with a voltage channel (of 31)

  • Personalized thresholds vs Standard thresholds: P = 0.003 — deceleration zones (DZs) as border zone tissue
  • Personalized thresholds vs Standard thresholds: P = 0.005 — VT isthmuses correlated with voltage channel

Further findings

Adjusted cutoff bipolar voltages best matching CMR images

Adjusted lower cutoff
0.51 ± 0.32 mV
Adjusted upper cutoff
1.79 ± 0.71 mV
High interindividual variability
from 0.14-1.68 to 0.7-3.21 mV

Original publication

Personalized voltage maps guided by cardiac magnetic resonance in the era of high-density mapping. Heart Rhythm. Published 25 April 2024. doi:10.1016/j.hrthm.2024.04.074

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.