Arrhythmias & electrophysiology
Precise anatomical landmark annotation of the cardiac conduction system on cardiac computed tomography angiography
Authors' conclusion
The method meets its clinically accepted margins with millimeter precision and good-to-excellent reliability, and transfers from expert to trained readers with bias within margin.
Study context
- Population
- CTA studies annotated for cardiac conduction system landmarksCTA studies, annotated by the expert and 6 trained readers
- Participants
- 50 studies
- Comparison
- Trained readers relative to an expert reference standard
- Trained readersn = 6
- Outcome
- Repeatability and reproducibility
Methodological detailsrepeatability and reproducibility, 3 sessions, equivalence testing of bias relative to expert
- Repeatability and reproducibility
- 3 sessions
- Equivalence testing of bias relative to expert
Main finding
As reported in the abstract
Repeatability SD
1.67-3.98 mm
Reproducibility SD
2.03-5.60 mm
Repeat reads within margin
88.6-97.4%
Reader means within margin
71.4-94.9%
Key notes
- All mean deviations within margin
- Clinically accepted error margins (2-5 mm)
Further findings
Inter-observer intraclass correlations (ICC) on annular heights
- Atrioventricular points
- 0.76-0.82
- Atrial points
- 0.79-0.98
Shared bias relative to the expert
- Shared bias
- 0.14-3.24 mm
- Apparent separation that was repeatability noise at B3
- 95%
- Apparent separation that was repeatability noise at B2
- 83%
- Within margin for all landmarks
- Equivalent within half the margin for sinus node, Point B, B2 and B3
Original publication
Precise anatomical landmark annotation of the cardiac conduction system on cardiac computed tomography angiography. Heart Rhythm. Published 17 November 2026. doi:10. 1016/ j. hrthm. 2026. 11. 016
Composed by AI from the published abstract and checked against it. How Briefs are made →