Arrhythmias & electrophysiology
Remote interrogation and reprogramming of cardiac implantable electronic devices using a custom multivendor solution
Authors' conclusion
Remote interrogation and reprogramming of CIEDs are feasible and safe across disparate clinical contexts and distances.
Study context
- Population
- Patients with cardiac implantable electronic devices (CIEDs) (pacemakers, ICDs from Abbott, Biotronik, MicroPort) in clinically driven encounters, within hospital, medium range (50–100 km) and long range (>5000 km)115 remote interrogations
- Participants
- 110 patients
- Outcome
- Technical feasibility (% of success) and safety (absence of complication) of remote interrogation and reprogramming when indicated
Methodological detailsmulticentre, Teleworking in 39 cases
- Multicentre
- Teleworking in 39 cases
Main finding
As reported in the abstract
Reprogramming sessions
56 of 115
49%
Further findings
Time lag for screen interaction
- Time lag below 1 second
Safety
- No clinical or technical complications
Original publication
Remote interrogation and reprogramming of cardiac implantable electronic devices using a custom multivendor solution. Heart Rhythm. Published 13 December 2022. doi:10. 1016/ j. hrthm. 2022. 12. 015
Composed by AI from the published abstract and checked against it. How Briefs are made →