Arrhythmias & electrophysiology

Safety and Feasibility of 3D Intracardiac Echocardiography in Guiding Left Atrial Appendage Occlusion With WATCHMAN FLX

JACC AdvPublished prospective multicenter study

Authors' conclusion

The novel 3-dimensional ICE VeriSight Pro Probe can safely and effectively guide LAAO with Watchman FLX using a simplified imaging protocol.

Study context

Population
Patients undergoing left atrial appendage occlusion (LAAO) with the Watchman FLX device, guided by a novel 3-dimensional intracardiac echocardiography (ICE) catheter (VeriSight Pro, Philips) under moderate sedationpatients undergoing LAAO with the Watchman FLX device
Participants
100 patients
Outcome
Feasibility endpoint: successful implantation of the FLX device and adequate seal (peri-device leak <5 mm)
Follow-up
45 days

Main finding

As reported in the abstract

ICE-guided LAAO with Watchman FLX

95%

95 of 100

Not implanted due to unsuitable anatomy
4 patients
Leak >5 mm at 45 days
1 patient

Observational data: associations, not effects.

Further findings

Baseline characteristics

Mean age
77.0 ± 8.1 years
Women
39 (39%)
CHA2DS2-VASc score
4.8 ± 1.5
HAS-BLED score
3.2 ± 1.1

Procedural characteristics

Total procedural time
54 ± 25 minutes
Mean contrast volume
41 ± 46 ml
No contrast used
16 patients

Physicians' rating of overall ICE imaging quality (1 = lowest, 5 = highest)

Overall ICE imaging quality
4.6 ± 0.6

Safety

Major complications · 7 days

Periprocedural stroke
1
Gastrointestinal bleeds requiring transfusion
2
No other major complications

Original publication

Safety and Feasibility of 3D Intracardiac Echocardiography in Guiding Left Atrial Appendage Occlusion With WATCHMAN FLX. JACC Adv. Published 13 January 2025. doi:10.1016/j.jacadv.2024.101570

Summarised from the published abstract; every number is checked against it