Arrhythmias & electrophysiology

Safety and Feasibility of Pulsed-Field Ablation in Patients With Mechanical Prosthetic Valves

JACC Clin ElectrophysiolPublished Multicentre observational study

Authors' conclusion

PFA in patients with mechanical valves appears safe, but careful catheter handling is essential to avoid electromagnetic interference hindering PFA delivery.

Study context

Population
Patients with mitral and/or aortic mechanical valve replacement undergoing endocardial AF ablation with pulsed-field ablation (PFA)
Participants
30 patients
Outcome
Feasibility (and safety) of PFA in patients with mechanical prosthetic valves
Methodological detailsconsecutive patients, 6 institutions
  • Consecutive patients
  • 6 institutions

Main finding

As reported in the abstract

Mean procedural time

77.5 ± 29.9 minutes

Mean left atrial dwell time

43.1 ± 21.4 minutes

Mean fluoroscopy time

14.0 ± 6.4 minutes

Observational data: associations, not effects.

Further findings

Mean procedural time
77.5 ± 29.9 minutes
Mean left atrial dwell time
43.1 ± 21.4 minutes
Mean fluoroscopy time
14.0 ± 6.4 minutes

Safety

Patients with electromagnetic interference of the PFA catheter adjacent to the mitral prosthesis, without clinical implications
1
No periprocedural or 1-month postablation major or minor complications
No prosthetic valve dysfunction on postprocedural transthoracic echocardiography
No energy delivery to the mitral edge of the isthmus possible despite multiple catheter repositioning attempts

Original publication

Safety and Feasibility of Pulsed-Field Ablation in Patients With Mechanical Prosthetic Valves. JACC Clin Electrophysiol. Published 20 November 2024. doi:10.1016/j.jacep.2024.09.025

Composed by AI from the published abstract and checked against it. How Briefs are made →